Meet Inspiring Speakers and Experts at our 3000+ Global Conference Series Events with over 1000+ Conferences, 1000+ Symposiums
and 1000+ Workshops on Medical, Pharma, Engineering, Science, Technology and Business.

Explore and learn more about Conference Series : World's leading Event Organizer

Back

Thipaporn Jaroonsirimaneekul

Thipaporn Jaroonsirimaneekul

Khon Kaen University, Thailand

Title: Enumeration of residual white blood cells in leuko-depleted blood products

Biography

Biography: Thipaporn Jaroonsirimaneekul

Abstract

Introduction: Quality assurance of white blood cells (WBC) depletion require in QC blood products. Inline filter and automate centrifuge (Reveos-system) produce the leuko-depleted blood product in routine, therefore we need the new method for WBC counting, to assure the blood products.

Objective: To count WBC in leukocyte depleted red blood cell (LD-PRC) from inline filter or Reveos blood bag and leukocyte depleted pooled platelet concentrates (LD-PC).

Method: At least 1% were sampling from total units of LD-PRC (N=20) and LD-PC (N=12) in routine unit for WBC enumeration within 6-8 hours after collection. The WBC in samples was stain by fluorescent dye, use LED excitation and CCD detection technologies makes the WBC analysis result.

Result: Residual WBC in LD-PRC and LD-PC are 0.26+0.21 and 0.23+0.12x106 cells /unit (X+SD), respectively. The maximum WBC residual in both products are 0.47 and 0.35x106 cells/unit. The waiting time for results interpretation was almost 8 minutes per test.

Conclusion: The leuko-depleted blood product in blood transfusion centre, faculty of medicine, Khon Kaen University are accepted by standard of American Associated of Blood Transfusions (AABB) and European Standard, by the new standard of automatic residual leukocyte counting (rWBC-ADAM). And the technique was easy to use and effective in routine.